GOSSEN Mueller & Weigert 9
:
Moving-Coil Movement Moving-Iron Movement Bimetal Movement
Application
Measurement of direct current
or direct voltage
Precision measurement
of arithmetic mean value
Measurement of alternating current or
alternating voltage
Alternating current measurement
True RMS measurement True RMS measurement (TRMS)
With rectifier:
Measurement of alternating current or
alternating voltage
Measurement of rectified value, effective
value display with sinewave
The integrated slave pointer indicates the
highest attained value
Bearings
Rugged pivot bearings
with spring-loaded jewels
Rugged pivot bearings with spring-loaded jewels
Damping
Eddy-current damping Viscous damping Thermal, time-delayed, for display of mean
effective value
Overshoot 5% of scale length 5% of scale length
Response Time 2 s per DIN EN 60061-1 2 s per DIN EN 60061-1 15 min., alternatively 8 min.
Reference Conditions
Frequency With rectifier: 45 Hz ... 65 Hz 45 Hz ... 65 Hz 45 Hz ... 65 Hz
Nominal Range of Use
Frequency With rectifier:
Ammeter: 40 Hz ... 1000 Hz
Voltmeter: 40 Hz ... 10000 Hz
Ammeter: 15 Hz ... 400 Hz
Voltmeter: 15 Hz ... 100 Hz
400 Hz
Scale Characteristics
nearly linear Lower measuring range value is
approx. 10% of upper measuring range value.
Ammeters upon request with double overload scale
Measuring Range
With rectifier:
for connection to transformer
=120% of rated transformer value
For connection to transformer
=120% of rated transformer value,
Ammeters upon request,
=100% of rated transformer value
For connection to transformer
=120% of rated transformer value
Overload Capacity
Continuous 120% of rated value 120% of rated value 120% of rated value
Short-Term:
Current Measurement 10 x rated value, 5 s 10 x rated value, 5 s 10 x rated value, 1 s (I
max
=10 A)
With rectifier: 40 x rated value, 1 s
2 x rated value, 0.5 s I
max
=250 A
Voltage Measurement 2 x rated value, 5 s 2 x rated value, 5 s
With rectifier:
2 x rated value, 0.5 s
Connection
Ammeters 40 A adjusted with horizontal cables to
the outside
Intrinsic Consumption
Ammeters: 0.4 ... 0.6 VA For rated transformer current:
1 A: approx. 1.6 VA
(approx. 1.1 VA for BM 48)
5 A: approx. 2.5 VA
(approx. 1.9 VA for BM 48)
Voltmeters: approx. 4.0 VA
See individual technical data for technical descriptions of frequency meters, active and reactive power meters and power factor meters.
Measuring Systems
Measuring Systems Technical Descriptions