Calibration Functions on Standard Balances
Select the appropriate menu code to define the access status for each
of the calibration functions, which are activated by holding down the t key for
a few seconds.
If the menu access switch is unlocked (“accessible” status indicated by -C- after
you have accessed the balance operating menu), the “external calibration”
function will be accessible even though you have set the menu code 1 9 2 for
“access denied.
External calibration Code
Accessible 1 9 1*
Access denied 1 9 2
Internal calibration Code
Accessible 1 10 1*
Access denied 1 10 2
Calibration test Code
Accessible 1 11 1*
Access denied 1 11 2
External Linearization
The linearization weights to be loaded are displayed on the balance one
after the other in increasing order. The balance must be re-calibrated after
external linearization.
Important Note:
If you use linearization weights that differ from the values displayed, this will
cause errors.
External linearization Code
Accessible 1 12 1
Access denied 1 12 2*
Internal linearization Code
Accessible 1 13 1*
Access denied 1 13 2
* = factory setting
2–14