Specifications
for Verified Balances
with an EC Type Approval
Model MC 210 S MC 210 P MC 410 S
-0CE -0CE -0CE
Type KB BA 100 KB BA 100 KB BA 100
Accuracy class kk k
Weighing range structure SuperRange PolyRange SuperRange
Maximum capacity Max* g 210 210 410
Scale interval d* mg 0.01 0.01/0.02/0.05 0.1
Verification scale interval e* mg 1 1 1
Minimum capacity Min* mg 1 1 10
Tare range (subtractive) % ––––––––––––––– ≤105 of Max ––––––––––––––
Range of use according to
Directive No. 90/384/EEC* g 0.001–210 0.001–210 0.001–410
Response time (average) s ––––––––––––– 10 ––––––––––––––––– 2.5 –––
Adaptation to ambient conditions
and application requirements s ––– By selection of 1 of 4 optimized filter levels ––
Display update (dep. on the filter level select.) –––––––––––––––– 0.2–0.4 ––––––––––––––––
Allowable ambient operating temperature
°C –––––––––––––––– +5...+40 ––––––––––––––––
Operating temperature range °C ––––––––––––––– +10 ... +40 ––––––––––––––
Pan size mm –––––––––––––––––– Ø 90––––––––––––––––––
Weighing chamber approx., (Ø x H) mm –––––––––––––––– 167 x 245 –––––––––––––––
Net weight, approx. kg –––––––––––––––––––12 –––––––––––––––––––
Power requirements V~ TNG6 230 VAC or 115 VAC, –20%...+15%
Frequency Hz ––––––––––––––––– 48–60 ––––––––––––––––
Power consumption approx. VA ––––––––– maximum: 28; average: 16 ––––––––
Selectable weight units g, mg
Selectable application programs Mass unit conversion; “EUREKA” air buoyancy
correction progr.; diff. weighing and
backweighing;
density determination; tare memory; net total;
weighing in percent; over/under checkweighing
(classification and filling); counting
Automatic zero tracking function
(can be turned off by menu code) –––––––––––––––– Standard ––––––––––––––––
Built-in interface
RS-232 C-S/V24-V28, RS-423/V 10; 7-bit;
parity: even, odd; transmission rates:
150...19,200
baud; 1or 2 stop bits;
software/hardware handshake
Standard features/equipment supplied
Dust cover/power supply xx x
Hanger for below-balance weighing x x x
Self-calibrating function “isoCAL” x x x
ISO/GLP recording capability/numeric keys x x x
Numeric keys x x x
* = EC Directive No. 90/384/EEC for non-automatic weighing instruments applicable in the
EC/EU Member States