TM 11-6625-2611-12
CHAPTER 5
ORGANIZATIONAL MAINTENANCE INSTRUCTIONS
5-1. General
Maintenance of the test set consists of cleaning the ex-
terior, replacing lamps and fuses, and minor cable
repair.
5-2. Cleaning
WARNING
The fumes of TRICHLOROETHANE are tox-
ic. Provide thorough ventilation whenever it is
used; avoid prolonged or repeated breathing
of vapor. Do not use near an open flame or hot
surface; trichloroethane is nonflammable but
heat converts the fumes to a highly toxic
phosgene gas the inhalation of which could
result in serious injury or death. Prolonged or
repeated skin contact with trichloroethane
can cause skin inflammation. When neces-
sary, use gloves, sleeves and aprons which the
solvent cannot penetrate.
CAUTION
Do not use trichloroethane on any painted or
silk-screened surface of the test set.
Clean the front panel and control knobs with a soft,
clean cloth. If dirt is difficult to remove, dampen the
cloth with water and use a mild soap. Clean exterior
metal surfaces as follows:
a. Remove moisture and loose dirt with a clean cloth
(item 2, app E).
b. Remove grease, fungus, and ground-in-dirt with
isopropyl alcohol (item 6, app E).
c. Clean unpainted surfaces with trichloroethane
(item 4, app E).
d. Remove dirt from hard-to-reach areas with a
brush (item 3, app E).
5-3. Touchup Painting Instructions
Remove rust and corrosion from metal surfaces by
lightly sanding with fine sandpaper (item 1, app E).
Brush two thin coats of paint on the bare metal to pro-
tect from further corrosion. Refer to applicable clean-
ing and refinishing practices specified in TB746-10.
5-4. Replacement of POWER Indicator
Lamp
Remove and replace the lamp as follows:
a. Unscrew indicator cap by turning it counterclock-
wise.
b. Press lamp in and turn it counterclockwise to
remove.
C. Install a replacement lamp (NE51H, NSN 6240-
00-683-3411 ) into lamp socket.
d. Tighten indicator cap by turning it clockwise.
5-5. Replacement of POWER DC FAULT in-
dicator Lamp
Remove and replace the lamp as follows:
a. Unscrew indicator cap by turning it counterclock-
wise.
b. Remove lamp from indicator cap.
c. Install replacement lamp (MS25237-327, NSN
6240-00-155-7836) into indicator cap.
d. Tighten indicator cap by turning it clockwise.
5-6. Replacement of Fuse
Remove and replace the fuse as follows:
a Unscrew fuse cap by turning it counterclockwise.
b. Pull defective fuse out of fuse cap and insert a
replacement fuse.
c. Reinstall replacement fuse (F02B250V1A, NSN
5920-00-284-9220) and fuse cap and tighten by turn-
ning clockwise.
5-7. Minor Repair of Cables
Repair minor cuts in cable insulation by covering it
first with insulation tape, and then with plastic tape. If
a cable is broken, ship defective cable to higher categw
ry of maintenance for repair.
5-8. CHAL AUX MOD DLY Knob Adjustment
If delay between main rf output and auxiliary output is
not 0.2
µ sec when control is set to detent, proceed as
follows:
a Connect test set and oscilloscope to 115 vac power
source.
b. Connect test set up as shown in figure 2-3 and set
oscilloscope controls as described in table 2-2 except as
follows:
A VOLTS/DIV
0.2
EXT 10/EXT/lNT/LINE
EXT
MAIN TIME/DIV
.1 µ SEC
C. Adjust test set controls as described in table 2-3
except as follows
MEASUREMENT FUNCTION PRF CHAL
SELECT switch
SCOPE TRIG/FREQ MEAS
Adjust to locate first challenge
DELAY (USEC) controls
pulse
d. Adjust SCOPE TRIG/FREQ MEAS DELAY
(USEC) MULT control to reference leading edge of
first pulse at center of Oscilloscope graticule.
e. Connect RF IN/OUT AUX jack to LOW PWR
MEAS IN jack.
f. Adjust CHAL AUX MOD DLY control to detent
position.
g. Loosen setscrews and hold knob so that ball
plunger remains seated in detent.
5-1