Load and Prime Reagents
HiSeq 2500 System User Guide
53
4 Slide the reagent rack out of the reagent compartment using the rack handle.
5 Remove the caps from each reagent tube and place the tube onto the rack in the
associated numbered position or matching label color.
Position
Reagent Description
17
HP8 or HP12 Index 1 (i7) Sequencing Primer Mix
18
HP3 Denaturation Solution
19
HT2 Wash Buffer
Table 6 Single-Indexed Run on a Single-Read Flow Cell or Paired-End Flow Cell
Position
Reagent Description
16
HP9 Index 2 (i5) SR Sequencing Primer Mix
17
HP8 or HP12 Index 1 (i7) Sequencing Primer Mix
18
HP3 Denaturation Solution
19
HT2 Wash Buffer
Table 7 Dual-Indexed Run on a Single-Read Flow Cell
Position
Reagent Description
10
RMR Resynthesis Mix
17
HP8 or HP12 Index 1 (i7) Sequencing Primer Mix
18
HP3 Denaturation Solution
19
HT2 Wash Buffer
Table 8 Dual-Indexed Run on a Paired-End Flow Cell
6 Place 15 ml conical tubes filled with 10 ml laboratory-grade water in unused rack
positions.
7 Slide the reagent rack into the reagent compartment, aligning the rack with the raised
guide on the floor of the compartment.
8 Lower the sippers into the tubes on the paired-end reagent rack as follows:
a Pull the handle towards you and then lower the handle while pulling it towards
you.
b Visually inspect the sippers to make sure that they do not bend as they lower into
the tubes.
c Release the handle into the slot on the bottom end of the groove.
9 Close the reagent compartment door.