Podsumowanie treści

Strona 1 - Nanoflow Pump

Agilent TechnologiesAgilent 1260 Infinity Nanoflow PumpUser Manual

Strona 2

10 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpIntroduction to the PumpHydraulic Path OverviewThe NanoFlow Pump is

Strona 3 - Contents

100 1260 Nanoflow Pump User Manual7 Error InformationModule Error MessagesTemperature Out of Range Temperature Out of Range 0: Pump channel ATemperatu

Strona 4

1260 Nanoflow Pump User Manual 101Error Information7Module Error MessagesMotor-Drive PowerThe current drawn by the pump motor exceeded the maximum lim

Strona 5

102 1260 Nanoflow Pump User Manual7 Error InformationModule Error MessagesEncoder MissingThe optical encoder on the pump motor in the module is missin

Strona 6

1260 Nanoflow Pump User Manual 103Error Information7Module Error MessagesElectro-Magnetic-Proportional-Valve (EMPV) MissingEMPV MissingThe EMPV in the

Strona 7

104 1260 Nanoflow Pump User Manual7 Error InformationModule Error MessagesServo Restart FailedThe pump motor in the module was unable to move into the

Strona 8

1260 Nanoflow Pump User Manual 105Error Information7Module Error MessagesPump Head MissingThe pump-head end stop in the pump was not found.When the pu

Strona 9 - Figure 1 Overview of the Pump

106 1260 Nanoflow Pump User Manual7 Error InformationModule Error MessagesIndex AdjustmentThe encoder index position in the module is out of adjustmen

Strona 10 - Hydraulic Path Overview

1260 Nanoflow Pump User Manual 107Error Information7Module Error MessagesStroke LengthThe distance between the lower piston position and the upper mec

Strona 11 - Introduction to the Pump

108 1260 Nanoflow Pump User Manual7 Error InformationModule Error MessagesWait TimeoutWhen running certain tests in the diagnostics mode or other spec

Strona 12

1091260 Nanoflow Pump User ManualAgilent Technologies8Test Functions and CalibrationMicro Mode Pressure Test 110Running the Test from the Agilent Lab

Strona 13

1260 Nanoflow Pump User Manual 11Introduction to the 1260 Infinity Nanoflow Pump1Introduction to the PumpFigure 2 The Hydraulic Path of the NanoFlow P

Strona 14

110 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationMicro Mode Pressure TestMicro Mode Pressure TestDescriptionThis is a fast test to ve

Strona 15

1260 Nanoflow Pump User Manual 111Test Functions and Calibration8Micro Mode Pressure TestMicro Mode Pressure Test ResultsThe test results are evaluate

Strona 16 - Early Maintenance Feetback

112 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationNormal Mode Pressure TestNormal Mode Pressure TestDescriptionThe System Pressure Tes

Strona 17 - Instrument Layout

1260 Nanoflow Pump User Manual 113Test Functions and Calibration8Normal Mode Pressure TestPositioning the Blank NutIf a specific component is suspecte

Strona 18 - Electrical Connections

114 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationNormal Mode Pressure TestRunning the Pressure TestRunning the test from the Agilent

Strona 19 - Rear View of the Module

1260 Nanoflow Pump User Manual 115Test Functions and Calibration8Normal Mode Pressure TestEvaluating the ResultsThe sum of all leaks between the pump

Strona 20 - Interfaces

116 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationLeak TestLeak TestLeak Test DescriptionThe Leak Test is a built-in troubleshooting t

Strona 21

1260 Nanoflow Pump User Manual 117Test Functions and Calibration8Leak TestPlateau 2Piston B2 delivers with a flow rate of 3 µL/min for 30 s.Ramp 5Pist

Strona 22 - Overview Interfaces

118 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationLeak TestRunning the Leak TestRunning the test from the Agilent Lab Advisor software

Strona 23 - Analog Signal Output

1260 Nanoflow Pump User Manual 119Test Functions and Calibration8Leak TestEvaluating the ResultsDefective or leaky components in the pump head lead to

Strona 24 - APG Remote

12 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpIntroduction to the PumpHow Does the Pumping Unit Work?Both pumping

Strona 25 - Special Interfaces

120 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationLeak TestPressure limit not reached but plateaus horizontal or positiveAll plateaus

Strona 26

1260 Nanoflow Pump User Manual 121Test Functions and Calibration8Leak TestFirst plateau negative or unstable, and at least one other plateau positiveS

Strona 27

122 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationLeak TestThird plateau negative or unstable and at least one other plateau positiveP

Strona 28

1260 Nanoflow Pump User Manual 123Test Functions and Calibration8Leak TestFourth plateau negative or unstable and at least one other plateau positiveP

Strona 29

124 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationEMPV TestEMPV TestEMPV Test DescriptionThe test is designed to verify the performanc

Strona 30

1260 Nanoflow Pump User Manual 125Test Functions and Calibration8EMPV CleaningEMPV Cleaning1260 Nanoflow Pump EMPV Cleaning DescriptionDepending on th

Strona 31 - Special Settings

126 1260 Nanoflow Pump User Manual8 Test Functions and CalibrationEMPV Cleaning

Strona 32 - Forced Cold Start

1271260 Nanoflow Pump User ManualAgilent Technologies9MaintenanceIntroduction to Maintenance 128Warnings and Cautions 129Overview of Maintenance 130Cl

Strona 33

128 1260 Nanoflow Pump User Manual9 MaintenanceIntroduction to MaintenanceIntroduction to MaintenanceThe pump is designed for easy repair. The most fr

Strona 34 - Site Requirements

1260 Nanoflow Pump User Manual 129Maintenance9Warnings and CautionsWarnings and CautionsWARNINGToxic, flammable and hazardous solvents, samples and re

Strona 35 - Power Cords

1260 Nanoflow Pump User Manual 13Introduction to the 1260 Infinity Nanoflow Pump1Introduction to the Pumpcloses the outlet valve, preventing any chamb

Strona 36 - Condensation

130 1260 Nanoflow Pump User Manual9 MaintenanceOverview of MaintenanceOverview of MaintenanceFigure 17 on page 131 shows the main assemblies of the pu

Strona 37 - Physical Specifications

1260 Nanoflow Pump User Manual 131Maintenance9Overview of MaintenanceFigure 17 Overview of Repair Procedures&'()*+,-.&%&'2 Activ

Strona 38 - Performance Specifications

132 1260 Nanoflow Pump User Manual9 MaintenanceCleaning the ModuleCleaning the ModuleThe module case should be kept clean. Cleaning should be done wit

Strona 39

1260 Nanoflow Pump User Manual 133Maintenance9Early Maintenance Feedback (EMF)Early Maintenance Feedback (EMF)Maintenance requires the exchange of com

Strona 40

134 1260 Nanoflow Pump User Manual9 MaintenanceEarly Maintenance Feedback (EMF)EMF CountersThe pump provides a series of EMF counters for the pump hea

Strona 41 - Installing the Module

1260 Nanoflow Pump User Manual 135Maintenance9Early Maintenance Feedback (EMF)Using the EMF CountersThe user-settable EMF limits for the EMF counters

Strona 42 - Unpacking the Module

136 1260 Nanoflow Pump User Manual9 MaintenanceChecking and Cleaning the Solvent Inlet FiltersChecking and Cleaning the Solvent Inlet FiltersWhen If s

Strona 43

1260 Nanoflow Pump User Manual 137Maintenance9Checking and Cleaning the Solvent Inlet FiltersCleaning the Solvent Filters1 Remove the blocked solvent

Strona 44 - One Stack Configuration

138 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Active Inlet Valve Cartridge or the Active Inlet ValveExchanging the Active Inlet Valve

Strona 45

1260 Nanoflow Pump User Manual 139Maintenance9Exchanging the Active Inlet Valve Cartridge or the Active Inlet ValveExchanging the Valve Cartridge1 Usi

Strona 46 - 3 Installing the Module

14 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpIntroduction to the PumpFor pump specifications, see “Performance Sp

Strona 47 - Two Stack Configuration

140 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Active Inlet Valve Cartridge or the Active Inlet ValveReplacing the Active Inlet Valve B

Strona 48

1260 Nanoflow Pump User Manual 141Maintenance9Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve6 After an exchange of the valve ca

Strona 49 - Installing the Pump

142 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Outlet Valve SieveExchanging the Outlet Valve Sieve1 Using a 1/4 inch wrench disconnect

Strona 50

1260 Nanoflow Pump User Manual 143Maintenance9Exchanging the Outlet Valve Sieve6 Check that the new valve is assembled correctly and that the gold sea

Strona 51

144 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Solvent Selection ValveExchanging the Solvent Selection ValveWhen If leaking internally,

Strona 52 - Connecting a Vacuum Degasser

1260 Nanoflow Pump User Manual 145Maintenance9Exchanging the Solvent Selection Valve1 Disconnect the solvent tubes and the active inlet valve connecti

Strona 53

146 1260 Nanoflow Pump User Manual9 MaintenanceRemoving and Disassembling the Pump HeadRemoving and Disassembling the Pump HeadWhen • Exchanging pump

Strona 54 - Flow Connections

1260 Nanoflow Pump User Manual 147Maintenance9Removing and Disassembling the Pump Head3Place the pump head on a flat surface. Loosen the lock screw (t

Strona 55

148 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Pump Seals and Seal Wear-in ProcedureExchanging the Pump Seals and Seal Wear-in Procedur

Strona 56

1260 Nanoflow Pump User Manual 149Maintenance9Exchanging the Pump Seals and Seal Wear-in Procedure3Clean the pump chambers with lint free cloth. Ensur

Strona 57

1260 Nanoflow Pump User Manual 15Introduction to the 1260 Infinity Nanoflow Pump1Introduction to the PumpHow Does Variable Stroke Volume Work?Due to t

Strona 58

150 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Pump Seals and Seal Wear-in ProcedureSeal Wear-in Procedure1 Place a bottle with 100 ml

Strona 59

1260 Nanoflow Pump User Manual 151Maintenance9Exchanging the PistonsExchanging the PistonsWhen When scratchedTools required • 3-mm hexagonal key• 4-mm

Strona 60

152 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Pistons3Reassemble the pump head assembly (see “Reassembling the Pump Head Assembly” on

Strona 61 - Using the Pump

1260 Nanoflow Pump User Manual 153Maintenance9Exchanging the Flow SensorExchanging the Flow Sensor1 Turn off the pump.2 Using a 1/4 inch wrench discon

Strona 62 - Pump Issues

154 1260 Nanoflow Pump User Manual9 MaintenanceReassembling the Pump Head AssemblyReassembling the Pump Head AssemblyTools required • 3-mm hexagonal k

Strona 63 - Fused Silica Capillary Issues

1260 Nanoflow Pump User Manual 155Maintenance9Reassembling the Pump Head Assembly3Carefully insert the pistons into the pump head assembly and press t

Strona 64

156 1260 Nanoflow Pump User Manual9 MaintenanceExchanging the Optional Interface BoardExchanging the Optional Interface Board1 Switch OFF the module a

Strona 65 - Algae Growth in HPLC Systems

1260 Nanoflow Pump User Manual 157Maintenance9Replacing Module FirmwareReplacing Module FirmwareTo upgrade/downgrade the module’s firmware carry out t

Strona 66

158 1260 Nanoflow Pump User Manual9 MaintenanceReplacing Module Firmware

Strona 67

1591260 Nanoflow Pump User ManualAgilent Technologies10Parts and Materials for MaintenancePump Housing and Main Assemblies 160Solvent Cabinet and Bott

Strona 68 - Method Parameters

16 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpEarly Maintenance FeetbackEarly Maintenance FeetbackMaintenance requ

Strona 69

160 1260 Nanoflow Pump User Manual10 Parts and Materials for MaintenancePump Housing and Main AssembliesPump Housing and Main AssembliesRepair Parts —

Strona 70 - Test Results and Evaluation

1260 Nanoflow Pump User Manual 161Parts and Materials for Maintenance10Pump Housing and Main AssembliesFigure 24 Overview of Main Assemblies (Front Vi

Strona 71 - Optimizing Performance

162 1260 Nanoflow Pump User Manual10 Parts and Materials for MaintenancePump Housing and Main AssembliesRepair Parts—Pump Housing and Main Assemblies

Strona 72 - 5 Optimizing Performance

1260 Nanoflow Pump User Manual 163Parts and Materials for Maintenance10Solvent Cabinet and Bottle-Head AssemblySolvent Cabinet and Bottle-Head Assembl

Strona 73 - Choosing the Right Pump Seals

164 1260 Nanoflow Pump User Manual10 Parts and Materials for MaintenanceHydraulic PathHydraulic PathItem p/n Description1 G1311-60003 Bottle-head asse

Strona 74

1260 Nanoflow Pump User Manual 165Parts and Materials for Maintenance10Hydraulic PathFigure 27 Hydraulic Path(''*(-,&&&'.&a

Strona 75 - The Standard Filter

166 1260 Nanoflow Pump User Manual10 Parts and Materials for MaintenancePump-Head AssemblyPump-Head AssemblyThe complete pump head assembly includes i

Strona 76

1260 Nanoflow Pump User Manual 167Parts and Materials for Maintenance10Pump-Head AssemblyFigure 28 Pump-Head Assembly&'()*+,-.&%&&

Strona 77

168 1260 Nanoflow Pump User Manual10 Parts and Materials for MaintenanceFlow Sensor AssemblyFlow Sensor AssemblyFigure 29 Flow Sensor Assemblyp/n Desc

Strona 78

1260 Nanoflow Pump User Manual 169Parts and Materials for Maintenance10Nanoflow Pump Accessory KitNanoflow Pump Accessory KitAccessory Kit (Nano Pump)

Strona 79

1260 Nanoflow Pump User Manual 17Introduction to the 1260 Infinity Nanoflow Pump1Instrument LayoutInstrument LayoutThe industrial design of the module

Strona 80 - System Pressure Test

170 1260 Nanoflow Pump User Manual10 Parts and Materials for MaintenanceNanoflow Pump Accessory Kit

Strona 81 - Leak Rate Test

1711260 Nanoflow Pump User ManualAgilent Technologies11Identifying CablesCable Overview 172Analog Cables 174Remote Cables 176BCD Cables 179CAN/LAN Cab

Strona 82

172 1260 Nanoflow Pump User Manual11 Identifying CablesCable OverviewCable OverviewAnalog cablesRemote cablesBCD cablesNOTENever use cables other than

Strona 83 - Module Status Indicator

1260 Nanoflow Pump User Manual 173Identifying Cables11Cable OverviewCAN cablesLAN cablesExternal Contact CableRS-232 cablesp/n Description5181-1516 CA

Strona 84 - User Interfaces

174 1260 Nanoflow Pump User Manual11 Identifying CablesAnalog CablesAnalog CablesOne end of these cables provides a BNC connector to be connected to A

Strona 85 - Agilent Lab Advisor Software

1260 Nanoflow Pump User Manual 175Identifying Cables11Analog CablesAgilent Module to BNC ConnectorAgilent Module to General Purposep/n 8120-1840 Pin B

Strona 86

176 1260 Nanoflow Pump User Manual11 Identifying CablesRemote CablesRemote CablesOne end of these cables provides a Agilent Technologies APG (Analytic

Strona 87 - Error Information

1260 Nanoflow Pump User Manual 177Identifying Cables11Remote CablesAgilent Module to 3396 Series III / 3395B IntegratorsAgilent Module to Agilent 3590

Strona 88 - 7 Error Information

178 1260 Nanoflow Pump User Manual11 Identifying CablesRemote CablesAgilent Module to General Purposep/n 01046-60201 Pin Universal Pin Agilent moduleS

Strona 89 - What Are Error Messages

1260 Nanoflow Pump User Manual 179Identifying Cables11BCD CablesBCD CablesOne end of these cables provides a 15-pin BCD connector to be connected to t

Strona 90 - General Error Messages

18 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpElectrical ConnectionsElectrical Connections• The CAN bus is a seria

Strona 91 - Synchronization Lost

180 1260 Nanoflow Pump User Manual11 Identifying CablesBCD CablesAgilent Module to 3396 Integratorsp/n 03396-60560 Pin 3396 Pin Agilent moduleSignal N

Strona 92 - Leak Sensor Open

1260 Nanoflow Pump User Manual 181Identifying Cables11CAN/LAN CablesCAN/LAN CablesBoth ends of this cable provide a modular plug to be connected to Ag

Strona 93 - Compensation Sensor Short

182 1260 Nanoflow Pump User Manual11 Identifying CablesExternal Contact CableExternal Contact CableOne end of this cable provides a 15-pin plug to be

Strona 94 - Fan Failed

1260 Nanoflow Pump User Manual 183Identifying Cables11Agilent Module to PCAgilent Module to PCp/n DescriptionG1530-60600 RS-232 cable, 2 mRS232-61600

Strona 95 - Module Error Messages

184 1260 Nanoflow Pump User Manual11 Identifying CablesAgilent 1200 Module to PrinterAgilent 1200 Module to Printerp/n Description5181-1529 Cable Prin

Strona 96 - Pressure Signal Missing

1851260 Nanoflow Pump User ManualAgilent Technologies12AppendixGeneral Safety Information 186The Waste Electrical and Electronic Equipment Directive 1

Strona 97 - Valve Failed

186 1260 Nanoflow Pump User Manual12 AppendixGeneral Safety InformationGeneral Safety Information The following general safety precautions must be obs

Strona 98 - Pump Configuration

1260 Nanoflow Pump User Manual 187Appendix12General Safety InformationMake sure that only fuses with the required rated current and of the specified t

Strona 99 - Inlet-Valve Fuse

188 1260 Nanoflow Pump User Manual12 AppendixGeneral Safety InformationSafety SymbolsTabl e 24 Safety SymbolsSymbol DescriptionThe apparatus is marked

Strona 100 - Temperature Limit Exceeded

1260 Nanoflow Pump User Manual 189Appendix12The Waste Electrical and Electronic Equipment DirectiveThe Waste Electrical and Electronic Equipment Direc

Strona 101 - Motor-Drive Power

1260 Nanoflow Pump User Manual 19Introduction to the 1260 Infinity Nanoflow Pump1Electrical ConnectionsRear View of the ModuleFigure 4 Rear View of th

Strona 102 - Inlet-Valve Missing

190 1260 Nanoflow Pump User Manual12 AppendixBatteries InformationBatteries InformationWARNINGLithium batteries may not be disposed-off into the domes

Strona 103 - Flow Sensor Missing

1260 Nanoflow Pump User Manual 191Appendix12Radio InterferenceRadio InterferenceCables supplied by Agilent Technologies are screened to provide optimi

Strona 104 - Servo Restart Failed

192 1260 Nanoflow Pump User Manual12 AppendixSound EmissionSound EmissionManufacturer’s DeclarationThis statement is provided to comply with the requi

Strona 105 - Index Limit

1260 Nanoflow Pump User Manual 193Appendix12Solvent InformationSolvent InformationObserve the following recommendations on the use of solvents.• Brown

Strona 106 - Index Missing

194 1260 Nanoflow Pump User Manual12 AppendixAgilent Technologies on InternetAgilent Technologies on InternetFor the latest information on products an

Strona 107 - Initialization Failed

1260 Nanoflow Pump User Manual 195IndexIndex88-bit configuration switchon-board LAN 26without On-Board LAN 28Aactive inlet valve 130, 138, 138Agilen

Strona 108 - Electronic fuse of SSV

196 1260 Nanoflow Pump User ManualIndexEMPV test 124EMPV cleaning 81encoder missing 102error messagesfan failed 94error messagescompensation sensor op

Strona 109 - Agilent Technologies

1260 Nanoflow Pump User Manual 197Indexremote timeout 91missing pressure reading 98motor-drive power 101Nnon-operating altitude 37non-operating temper

Strona 110 - Micro Mode Pressure Test

198 1260 Nanoflow Pump User ManualIndexrear view 48status indicator 83stroke length 107stroke volume 15synchronization lost 91system setup and install

Strona 112 - Normal Mode Pressure Test

1260 Nanoflow Pump User ManualNotices© Agilent Technologies, Inc. 2007, 2008, 2010No part of this manual may be reproduced in any form or by any means

Strona 113 - Positioning the Blank Nut

20 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpInterfacesInterfacesThe Agilent 1200 Infinity Series modules provide

Strona 114 - Running the Pressure Test

© Agilent Technologies 2007, 2008, 2010Printed in Germany 06/2010*G2226-90012**G2226-90012*G2226-90012www.agilent.comAgilent TechnologiesIn This BookT

Strona 115 - Evaluating the Results

1260 Nanoflow Pump User Manual 21Introduction to the 1260 Infinity Nanoflow Pump1Interfaces• CAN connectors as interface to other modules• LAN connect

Strona 116 - Leak Test

22 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpInterfacesOverview Interfaces CANThe CAN is inter-module communicati

Strona 117

1260 Nanoflow Pump User Manual 23Introduction to the 1260 Infinity Nanoflow Pump1InterfacesThe RS-232C is designed as DCE (data communication equipmen

Strona 118 - Running the Leak Test

24 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpInterfacesAPG RemoteThe APG Remote connector may be used in combinat

Strona 119

1260 Nanoflow Pump User Manual 25Introduction to the 1260 Infinity Nanoflow Pump1InterfacesSpecial InterfacesSome modules have module specific interfa

Strona 120 - All plateaus negative

26 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpSetting the 8-bit Configuration Switch (On-Board LAN)Setting the 8-b

Strona 121

1260 Nanoflow Pump User Manual 27Introduction to the 1260 Infinity Nanoflow Pump1Setting the 8-bit Configuration Switch (On-Board LAN)Legend:0 (switch

Strona 122

28 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpSetting the 8-bit Configuration Switch (On-Board LAN)Setting the 8-b

Strona 123

1260 Nanoflow Pump User Manual 29Introduction to the 1260 Infinity Nanoflow Pump1Setting the 8-bit Configuration Switch (On-Board LAN)Communication Se

Strona 124 - EMPV Test

1260 Nanoflow Pump User Manual 3ContentsContents1 Introduction to the 1260 Infinity Nanoflow Pump 7Introduction to the Pump 8Early Maintenance Feetbac

Strona 125 - EMPV Cleaning

30 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpSetting the 8-bit Configuration Switch (On-Board LAN)One start bit a

Strona 126

1260 Nanoflow Pump User Manual 31Introduction to the 1260 Infinity Nanoflow Pump1Setting the 8-bit Configuration Switch (On-Board LAN)Special Settings

Strona 127 - Maintenance

32 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpSetting the 8-bit Configuration Switch (On-Board LAN)Forced Cold Sta

Strona 128 - Introduction to Maintenance

331260 Nanoflow Pump User ManualAgilent Technologies2Site Requirements and SpecificationsSite Requirements 34Physical Specifications 37Performance Spe

Strona 129 - Warnings and Cautions

34 1260 Nanoflow Pump User Manual2 Site Requirements and SpecificationsSite RequirementsSite RequirementsA suitable environment is important to ensure

Strona 130 - Overview of Maintenance

1260 Nanoflow Pump User Manual 35Site Requirements and Specifications2Site RequirementsPower CordsDifferent power cords are offered as options with th

Strona 131

36 1260 Nanoflow Pump User Manual2 Site Requirements and SpecificationsSite RequirementsBench SpaceThe module dimensions and weight (see Table 13 on p

Strona 132 - Cleaning the Module

1260 Nanoflow Pump User Manual 37Site Requirements and Specifications2Physical SpecificationsPhysical SpecificationsTabl e 13 Physical SpecificationsT

Strona 133

38 1260 Nanoflow Pump User Manual2 Site Requirements and SpecificationsPerformance SpecificationsPerformance SpecificationsTabl e 14 Performance Speci

Strona 134 - EMF Counters

1260 Nanoflow Pump User Manual 39Site Requirements and Specifications2Performance SpecificationsAnalog output For pressure monitoring, 2 mV/bar, one o

Strona 135 - Using the EMF Counters

4 1260 Nanoflow Pump User ManualContents6 Troubleshooting and Diagnostics 79Overview of the Module’s Indicators and Test Functions 80Status Indicators

Strona 136 - 9 Maintenance

40 1260 Nanoflow Pump User Manual2 Site Requirements and SpecificationsPerformance Specifications

Strona 137

411260 Nanoflow Pump User ManualAgilent Technologies3Installing the ModuleUnpacking the Module 42Delivery Checklist 42Nanoflow Pump Accessory Kit 43Op

Strona 138

42 1260 Nanoflow Pump User Manual3 Installing the ModuleUnpacking the ModuleUnpacking the ModuleIf the delivery packaging shows signs of external dama

Strona 139 - Tools required Wrench 14 mm

1260 Nanoflow Pump User Manual 43Installing the Module3Unpacking the ModuleNanoflow Pump Accessory KitAccessory Kit (Nano Pump) (p/n G2226-68755)1260

Strona 140

44 1260 Nanoflow Pump User Manual3 Installing the ModuleOptimizing the Stack ConfigurationOptimizing the Stack ConfigurationIf your module is part of

Strona 141 - HdakZci^caZiijWZ

1260 Nanoflow Pump User Manual 45Installing the Module3Optimizing the Stack ConfigurationFigure 8 Recommended Stack Configuration for 1260 (Front View

Strona 142

46 1260 Nanoflow Pump User Manual3 Installing the ModuleOptimizing the Stack ConfigurationFigure 9 Recommended Stack Configuration for 1260 (Rear View

Strona 143 - DjiaZiWVaakVakZ

1260 Nanoflow Pump User Manual 47Installing the Module3Optimizing the Stack ConfigurationTwo Stack ConfigurationTo avoid excessive height of the stack

Strona 144

48 1260 Nanoflow Pump User Manual3 Installing the ModuleOptimizing the Stack ConfigurationFigure 11 Recommended Two Stack Configuration for 1260 (Rear

Strona 145 - HdakZcihZaZXi^dc

1260 Nanoflow Pump User Manual 49Installing the Module3Installing the PumpInstalling the PumpFor other cables see text below and “Cable Overview” on p

Strona 146

1260 Nanoflow Pump User Manual 5Contents10 Parts and Materials for Maintenance 159Pump Housing and Main Assemblies 160Solvent Cabinet and Bottle-Head

Strona 147

50 1260 Nanoflow Pump User Manual3 Installing the ModuleInstalling the Pump1 Place the module on the bench in a horizontal position.2 Ensure the power

Strona 148

1260 Nanoflow Pump User Manual 51Installing the Module3Installing the Pump5 Connect the required interface cables to the rear of the pump, see “Connec

Strona 149 - 8]VbWZgh

52 1260 Nanoflow Pump User Manual3 Installing the ModuleConnecting Modules and Control SoftwareConnecting Modules and Control SoftwareConnecting Modul

Strona 150 - Seal Wear-in Procedure

1260 Nanoflow Pump User Manual 53Installing the Module3Connecting Modules and Control SoftwareConnecting Control Software and/or G4208 A Instant Pilot

Strona 151 - Exchanging the Pistons

54 1260 Nanoflow Pump User Manual3 Installing the ModuleFlow ConnectionsFlow ConnectionsParts required # p/n Description1 Other modules1 G1376-68755 A

Strona 152

1260 Nanoflow Pump User Manual 55Installing the Module3Flow Connections1 Remove the front cover by pressing the snap fasteners on both sides.Figure 14

Strona 153 - Exchanging the Flow Sensor

56 1260 Nanoflow Pump User Manual3 Installing the ModuleFlow Connections7 Purge your system before first use (see “Priming Your System With the Pump”

Strona 154

1260 Nanoflow Pump User Manual 57Installing the Module3Flow Connections1 Capillary EMPV to Nano Flow sensor (4 µL flow sensor) (p/n G1375-87321)2 Cap

Strona 155

58 1260 Nanoflow Pump User Manual3 Installing the ModuleGet the System Ready for the First InjectionGet the System Ready for the First InjectionWhen y

Strona 156

1260 Nanoflow Pump User Manual 59Installing the Module3Get the System Ready for the First Injection7 Repeat steps 4 to 6 for the three remaining solve

Strona 157 - Replacing Module Firmware

6 1260 Nanoflow Pump User ManualContents

Strona 158

60 1260 Nanoflow Pump User Manual3 Installing the ModuleGet the System Ready for the First Injection

Strona 159

611260 Nanoflow Pump User ManualAgilent Technologies4Using the PumpHints For Successfully Using the Pump 62Solvent Information 64Algae Growth in HPLC

Strona 160

62 1260 Nanoflow Pump User Manual4 Using the PumpHints For Successfully Using the PumpHints For Successfully Using the PumpPump Issues• Flush the pump

Strona 161

1260 Nanoflow Pump User Manual 63Using the Pump4Hints For Successfully Using the PumpFused Silica Capillary Issues• When connecting a capillary (espec

Strona 162 - Item p/n Description

64 1260 Nanoflow Pump User Manual4 Using the PumpSolvent InformationSolvent InformationAlways filter solvents through 0.4 µm filters, small particles

Strona 163

1260 Nanoflow Pump User Manual 65Using the Pump4Algae Growth in HPLC SystemsAlgae Growth in HPLC SystemsThe presence of algae in HPLC systems can caus

Strona 164 - Hydraulic Path

66 1260 Nanoflow Pump User Manual4 Using the PumpAlgae Growth in HPLC Systems• Algae growth may also be the possible source for failures of the ball v

Strona 165 - Figure 27 Hydraulic Path

1260 Nanoflow Pump User Manual 67Using the Pump4Checkout procedure for a G2229A Nano LC SystemCheckout procedure for a G2229A Nano LC SystemUse this p

Strona 166 - Pump-Head Assembly

68 1260 Nanoflow Pump User Manual4 Using the PumpCheckout procedure for a G2229A Nano LC System8 Pump 0.6 µL/min, micro mode, 70 % A (water) / 30 % B

Strona 167 - Figure 28 Pump-Head Assembly

1260 Nanoflow Pump User Manual 69Using the Pump4Checkout procedure for a G2229A Nano LC SystemMethod Parameters Micro High Performance Autosampler• In

Strona 168 - Flow Sensor Assembly

71260 Nanoflow Pump User ManualAgilent Technologies1Introduction to the 1260 Infinity Nanoflow PumpIntroduction to the Pump 8Hydraulic Path Overview 1

Strona 169 - Nanoflow Pump Accessory Kit

70 1260 Nanoflow Pump User Manual4 Using the PumpCheckout procedure for a G2229A Nano LC SystemTest Results and EvaluationTypical pressure in bypass m

Strona 170

711260 Nanoflow Pump User ManualAgilent Technologies5Optimizing PerformanceHints for the Micro Vacuum Degasser 72Choosing the Right Pump Seals 73How t

Strona 171 - Identifying Cables

72 1260 Nanoflow Pump User Manual5 Optimizing PerformanceHints for the Micro Vacuum DegasserHints for the Micro Vacuum DegasserIf you are using the va

Strona 172 - Cable Overview

1260 Nanoflow Pump User Manual 73Optimizing Performance5Choosing the Right Pump SealsChoosing the Right Pump SealsThe standard seal for the pump can b

Strona 173

74 1260 Nanoflow Pump User Manual5 Optimizing PerformanceHow to Choose the Primary FlowHow to Choose the Primary FlowThe primary flow can be set in th

Strona 174 - Analog Cables

1260 Nanoflow Pump User Manual 75Optimizing Performance5The Standard FilterThe Standard FilterThe standard filter has a volume of typically 100 µL. If

Strona 175

76 1260 Nanoflow Pump User Manual5 Optimizing PerformanceHow to Optimize the Compressibility Compensation SettingHow to Optimize the Compressibility C

Strona 176 - Remote Cables

1260 Nanoflow Pump User Manual 77Optimizing Performance5How to Optimize the Compressibility Compensation Setting5 Starting with a compressibility sett

Strona 177

78 1260 Nanoflow Pump User Manual5 Optimizing PerformanceHow to Optimize the Compressibility Compensation Setting

Strona 178 - 11 Identifying Cables

791260 Nanoflow Pump User ManualAgilent Technologies6Troubleshooting and DiagnosticsOverview of the Module’s Indicators and Test Functions 80Status In

Strona 179 - BCD Cables

8 1260 Nanoflow Pump User Manual1 Introduction to the 1260 Infinity Nanoflow PumpIntroduction to the PumpIntroduction to the PumpThe low flow pumps co

Strona 180

80 1260 Nanoflow Pump User Manual6 Troubleshooting and DiagnosticsOverview of the Module’s Indicators and Test FunctionsOverview of the Module’s Indic

Strona 181 - CAN/LAN Cables

1260 Nanoflow Pump User Manual 81Troubleshooting and Diagnostics6Overview of the Module’s Indicators and Test FunctionsLeak Rate TestThe Leak Rate Tes

Strona 182 - External Contact Cable

82 1260 Nanoflow Pump User Manual6 Troubleshooting and DiagnosticsStatus IndicatorsStatus IndicatorsTwo status indicators are located on the front of

Strona 183 - Agilent Module to PC

1260 Nanoflow Pump User Manual 83Troubleshooting and Diagnostics6Status IndicatorsModule Status IndicatorThe module status indicator indicates one of

Strona 184

84 1260 Nanoflow Pump User Manual6 Troubleshooting and DiagnosticsUser InterfacesUser InterfacesDepending on the User Interface, the available tests v

Strona 185 - Appendix

1260 Nanoflow Pump User Manual 85Troubleshooting and Diagnostics6Agilent Lab Advisor SoftwareAgilent Lab Advisor SoftwareThe Agilent Lab Advisor softw

Strona 186 - General Safety Information

86 1260 Nanoflow Pump User Manual6 Troubleshooting and DiagnosticsAgilent Lab Advisor Software

Strona 187

871260 Nanoflow Pump User ManualAgilent Technologies7Error InformationWhat Are Error Messages 89General Error Messages 90Timeout 90Shut-Down 90Remote

Strona 188 - Safety Symbols

88 1260 Nanoflow Pump User Manual7 Error InformationAgilent Lab Advisor SoftwareFlow Sensor Missing 103Leak Sensor Missing 103Servo Restart Failed 104

Strona 189 - Abstract

1260 Nanoflow Pump User Manual 89Error Information7What Are Error MessagesWhat Are Error MessagesError messages are displayed in the user interface wh

Strona 190 - Batteries Information

1260 Nanoflow Pump User Manual 9Introduction to the 1260 Infinity Nanoflow Pump1Introduction to the PumpFigure 1 Overview of the Pump8HBWdVgYEjbeYg^

Strona 191 - Radio Interference

90 1260 Nanoflow Pump User Manual7 Error InformationGeneral Error MessagesGeneral Error MessagesGeneral error messages are generic to all Agilent seri

Strona 192 - Sound Emission

1260 Nanoflow Pump User Manual 91Error Information7General Error MessagesRemote TimeoutA not-ready condition is still present on the remote input. Whe

Strona 193 - Solvent Information

92 1260 Nanoflow Pump User Manual7 Error InformationGeneral Error MessagesLeak Sensor ShortThe leak sensor in the module has failed (short circuit).Th

Strona 194 - 12 Appendix

1260 Nanoflow Pump User Manual 93Error Information7General Error MessagesCompensation Sensor OpenThe ambient-compensation sensor (NTC) on the main boa

Strona 195

94 1260 Nanoflow Pump User Manual7 Error InformationGeneral Error MessagesFan FailedThe cooling fan in the module has failed. The hall sensor on the f

Strona 196

1260 Nanoflow Pump User Manual 95Error Information7Module Error MessagesModule Error MessagesZero Solvent CounterThe error message is triggered if the

Strona 197

96 1260 Nanoflow Pump User Manual7 Error InformationModule Error MessagesPressure Below Lower LimitThe system pressure has fallen below the lower pres

Strona 198

1260 Nanoflow Pump User Manual 97Error Information7Module Error MessagesValve Failed Valve 0 Failed: valve A1Valve 1 Failed: valve A2Valve 2 Failed: v

Strona 199

98 1260 Nanoflow Pump User Manual7 Error InformationModule Error MessagesMissing Pressure ReadingThe pressure readings read by the pump ADC (analog-di

Strona 200 - *G2226-90012*

1260 Nanoflow Pump User Manual 99Error Information7Module Error MessagesValve FuseValve Fuse 0: Channels A1 and A2Valve Fuse 1: Channels B1 and B2One