1260 Nanoflow Pump User Manual 13
Introduction to the 1260 Infinity Nanoflow Pump
1
Introduction to the Pump
closes the outlet valve, preventing any chamber 2 solvent from back-streaming
into chamber 1. After a predefined piston 1 stroke length, the servo motor is
stopped, and the active inlet valve is closed. The pistons now reverse
directions. Piston 1 begins its delivery stroke (high pressure), and piston 2
begins its intake stroke. Piston 2 is moving at only half the speed of piston 1.
The outlet valve is forced open by the pressure generated by piston 1. Piston 1
begins to deliver the volume previously sucked into chamber 1. Because of the
2:1 speed ratio of the pistons, half of the solvent flow from chamber 1 is forced
out of the pump head, continuing into the pump hydraulic path. The other half
of the flow from chamber 1 simultaneously refills chamber 2.
When piston 1 has completed its delivery stroke, the pistons reverse direction,
and the cycle is repeated.
Figure 3 Operating Principle of the Pump Head
;gdbhdakZciWdiiaZ
Idb^m^c\X]VbWZg
HZVa
E^hidc
7VaahXgZlYg^kZ
Bdidgl^i]ZcXdYZg
<ZVg
>caZi
kVakZ
kVakZ
DjiaZi
Przeglądanie stron 12